Rheometrics Solids Analyzer (RSA II) Specifications

Operating Parameter

Specifications

Actuator

Frequency range
Amplitude range
Bandwidth
Response time
Strain Resolution
Maximum force
0.01 to 100 rad/sec (0.0016 to 16 Hz)
± 0.5 mm
> 125 rad/sec (20 Hz) at 0.5 mm
< 5 msec to 90% of final value
± 0.05   (0.00005 mm)
9.81 N (1000 g)

Transducer

Maximum response time
 Phase angle resolution
Measurement sensitivity
5 msec rise time to 90% of a 50g step
± 0.1° (0.00174 rad)
1 g

Controller

Temperature                         standard: range                                     optional:
Temperature ramp rate
Temperature stability
Ambient to +500°C
- 150°C to +600°C
0.1°  to 50°C/min
+0.5°C at thermal equilibrium